ISO 9001
CE Certified
Make In India
Technical Specifications
| Control Modes | Typically offer both air mode (controlling the incubator's air temperature) and baby mode (servo-control based on the infant's skin temperature via a sensor). |
| Range | Adjustable humidity levels, typically from 40% to 90% RH, to reduce evaporative heat loss. |
| Accuracy/Variability | High accuracy is critical, with typical resolutions of 0.1°C and accuracy within ±0.2°C to ±0.5°C. |
| Safety | Automatic safety cut-offs and alarms are mandatory to prevent overheating (e.g., at 40°C air temperature). |
| Mechanism | Achieved by passing air over a water reservoir, often with a dedicated heater for the water. |
| Input | Connection for external oxygen supply. |
| Concentration | Adjustable concentration range, often 21% to 60% or higher, with digital display and alarms. |
| Canopy | A clear, enclosed plastic hood, often double-walled to minimize heat and air loss. |
| Access | Features portholes and hinged doors for easy access during care without opening the entire unit. |
| Mobility | Typically mounted on a trolley with casters and brakes for movement within the hospital. |
| Alarms | Comprehensive audio and visual alarms for power failure, fan failure, sensor failure, temperature deviations, and low water levels. |
| Integrated Monitoring | Modern units may include integrated weighing scales, X-ray trays, pulse oximetry (SpO2) and other vital sign monitors, and IV pump integration. |
| Power | Operates on mains electricity (AC-powered), with transport incubators also featuring a substantial battery backup (e.g., 4 hours) for mobility |
Similar Products
View all
RADIANT BABY WARMER
Baby warmer (Infant Radiant Warmer) is a medical device providing controlled overhead infrared heat ...

PHOTOTHERAPY MACHINE
Phototherapy is a medical treatment that uses specific wavelengths of natural or artificial light to...

BABY CRADLE
A baby cradle is a small, specialized bed for an infant, often designed with a rocking or swinging m...